TMJ-9003 TMJ-9003 Ozone Aging Tester
- Description
Ozone is the main factor in the cracking of petrochemical products such as plastic rubber. The ozone aging tester can be used to test plastic and rubber products with static and dynamic tensile deformation, such as vulcanized rubber, thermoplastic rubber, and cable insulation; according to the specified time, without light, In the case of constant ozone concentration and constant temperature, expose the sample to the sealed air in the ozone room, and then observe the degree of cracks on the surface of the sample and the degree of changes in other properties.
- Feature
- The inner box is made of ozone-resistant SUS#316 stainless steel.
- The NOx-free ozone generator module can be switched in sections :
low-concentration UV or high, medium and low-concentration corona.
discharge type ozone generator modules can be selected. - Real-time ozone concentration and temperature and humidity monitoring.
- Professionally designed sample fixture.
- Advanced touch panel, easy to operate.
MODEL SPECIFICATION |
TMJ-9003A | TMJ-9003B |
Internal (WxDxH)mm | 500 x 500 x 600 | |
External (WxDxH)mm | 1500 x 1500 x 2000 | |
Temp. Range | -20℃~100℃ | |
HUM. Range | N |
30-95%RH @ 40℃ |
Ozone monitoring module | 0.25~10ppm / 10~100ppm / 100~500ppm Provide more than two levels of concentration correction and control settings |
|
No nitrogen oxide ozone gas generation module | UV ozone generation module : 0.25-10ppm @ CDA or O2 CD discharge generation module : 0.25~10ppm/10~100ppm / 100-500ppm @ O2 |
|
Temp./ Hum. Accuracy | 0.1℃ | 0.1℃ / 0.1%RH |
Temp. Stability | ±1.0℃ | |
Temp. Uniformity | ≤ 2.0℃ | |
Wind speed | Measurement at the air outlet : 0.3~0.8 m / s (adjustable) | |
Power | 220V / 380V 1Φ / 3Φ |
Note : Ozone emission is discharged after being absorbed by activated carbon.